这项工作介绍了一种使用离子迁移谱法(IMS)、顶空进样(HS)和Tenax吸附剂预分离(HS-Tenax IMS)直接检测酒精饮料中甲醇的新方法。采用基于掺杂气体的电离方法来克服反应物离子与甲醇峰的重叠。测试了六种不同的酮类掺杂剂,并选择2-庚酮作为最佳变体。Tenax GR柱用于HS和作为IMS检测前甲醇和乙醇的预分离方法

Rapid and Direct Determination of Methanol in Alcoholic Beverages by Ion Mobility Spectrometry
Vahideh Ilbeigi et al.  DOI: 10.1021/acs.analchem.5c03428

Abstract
This work introduces a new method for direct detection of methanol in alcoholic beverages, using ion mobility spectrometry (IMS), headspace sampling (HS), and Tenax adsorbent preseparation (HS-Tenax-IMS). An ionization method based on the dopant gases was employed to overcome the overlapping of the reactant ions with the methanol peak. Six different ketones dopants were tested, and 2-heptanone was selected as the optimal variant. Tenax-GR column was used for both HS and as a preseparation method of methanol and ethanol prior to IMS detection. The optimum sampling time of 15 min was achieved, and analysis time was less than 1 min. The linear range for detection of methanol in solution was from 0.08% to 1.0%, corresponding to 630–7900 mg/L (R2 = 0.996). The obtained limit of quantification (LOQ) of 630 mg/L and limit of detection (LOD) of 210 mg/L are sufficient for efficient methanol detection in the food industry.

这项工作介绍了一种使用离子迁移谱法(IMS)、顶空进样(HS)和Tenax吸附剂预分离(HS-Tenax IMS)直接检测酒精饮料中甲醇的新方法。采用基于掺杂气体的电离方法来克服反应物离子与甲醇峰的重叠。测试了六种不同的酮类掺杂剂,并选择2-庚酮作为最佳变体。Tenax GR柱用于HS和作为IMS检测前甲醇和乙醇的预分离方法。最佳采样时间为15分钟,分析时间小于1分钟。溶液中甲醇的检测线性范围为0.08%至1.0%,对应于630-7900 mg/L(R2=0.996)。所获得的定量限(LOQ)为630 mg/L,检测限(LOD)为210 mg/L,足以在食品工业中有效检测甲醇。

 

 